Computer-assisted scan protocol and reconstruction (CASPAR)reduction of image noise and patient dose

Jonathan Sperl, Dirk Bequé, Bernhard Claus, Bruno De Man, Bob Senzig, Martin Brokate

Publikation: Beitrag in FachzeitschriftArtikelBegutachtung

42 Zitate (Scopus)

Abstract

X-ray computed tomography is a powerful medical imaging device. It allows high-resolution 3-D visualization of the human body. However, one drawback is the health risk associated with ionizing radiation. Simply downscaling the radiation intensities over the entire scan results in increased quantum noise. This paper proposes the concept of computer-assisted scan protocol and reconstruction. More specifically, we propose a method to compute patient and task-specific intensity profiles that achieve an optimal tradeoff between radiation dose and image quality. Therefore, reasonable image variance and dose metrics are derived. Conventional third-generation systems as well as inverted geometry concepts are considered. Two dose/noise minimization problems are formulated and solved by an efficient algorithm providing optimized milliampere (mA)-profiles. Thorax phantom simulations demonstrate the promising advantage of this technique: in this particular example, the dose is reduced by 53% for third-generation systems and by 86% for an inverted geometry in comparison to a sinusoidal mA-profile at a constant upper noise limit.

OriginalspracheEnglisch
Aufsatznummer5423295
Seiten (von - bis)724-732
Seitenumfang9
FachzeitschriftIEEE Transactions on Medical Imaging
Jahrgang29
Ausgabenummer3
DOIs
PublikationsstatusVeröffentlicht - März 2010

Fingerprint

Untersuchen Sie die Forschungsthemen von „Computer-assisted scan protocol and reconstruction (CASPAR)reduction of image noise and patient dose“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren