Comparative study of the corrosion behavior of peripheral stents in an accelerated corrosion model: Experimental in vitro study of 28 metallic vascular endoprostheses

Karolin J. Paprottka, Philipp M. Paprottka, Maximilian F. Reiser, Tobias Waggershauser

Publikation: Beitrag in FachzeitschriftArtikelBegutachtung

13 Zitate (Scopus)

Abstract

PURPOSE Clinical cases of stent-fractures show that corrosion behavior might play a role in these fractures. Implanted in vivo, especially in combination with other implanted foreign materials, these metallic products are exposed to special conditions, which can cause a process of corrosion. Here, we aimed to test the corrosion potential of stents made of different materials in an in vitro setting. METHODS A total of 28 peripheral stents of different materials (nitinol, cobalt-chromium-nickel, tantalum, V4A) and surface treatments (electropolish, mechanical polish, no polish) were tested in vitro. Corrosion was accelerated by applying a constant voltage of 3.5 V and amperage of 1.16 mA in 0.9% NaCl. RESULTS Nitinol stents showed the lowest susceptibility to corrosion and the longest period without damage. The Memotherm II® (BARD Angiomed®) was the only stent that showed neither macroscopic nor microscopic damages. The worst performing material was cobalt-chromium-nickel, which showed corrosion damages about ten times earlier compared to nitinol. Considering the reasons for termination of the test, nitinol stents primarily showed length deficits, while V4A and tantalum stents showed fractures. Cobalt-chromium-nickel stents had multiple fractures or a completely sis in equal proportions. When placed in direct contact, nitinol stents showed best corrosion resistance, regardless of what material they were combined with. In terms of polishing treatments, electropolished stents performed the best, mechanical-polished stents and those without polishing treatment followed. CONCLUSION The analysis of corrosion behavior may be useful to select the right stent fulfilling the individual needs of the patient within a large number of different stents.

OriginalspracheEnglisch
Seiten (von - bis)403-409
Seitenumfang7
FachzeitschriftDiagnostic and Interventional Radiology
Jahrgang21
Ausgabenummer5
DOIs
PublikationsstatusVeröffentlicht - 1 Sept. 2015

Fingerprint

Untersuchen Sie die Forschungsthemen von „Comparative study of the corrosion behavior of peripheral stents in an accelerated corrosion model: Experimental in vitro study of 28 metallic vascular endoprostheses“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren