Abstract
Bioreactors with a controlled physiological environment are being developed to study various cell processes. The influences of mechanostimulation on bone cell cultures can be investigated using a compression bioreactor. The developed bioreactor system applies a cyclic compression force to the specimen via an eccentrically mounted push rod. The compression force is monitored by a force sensor to detect changes in the material properties of the specimen. Depending on the piston setting, a stroke of 0.28 - 2.50 mm can be applied to the specimen. The bioreactor system was tested with a trial run of 18 days. A sample was continuously stimulated with a loading frequency of 2 Hz and a stroke of 1.50 mm.
Originalsprache | Englisch |
---|---|
Seiten (von - bis) | 899-902 |
Seitenumfang | 4 |
Fachzeitschrift | Current Directions in Biomedical Engineering |
Jahrgang | 7 |
Ausgabenummer | 2 |
DOIs | |
Publikationsstatus | Veröffentlicht - 1 Okt. 2021 |